The Calibration of Microwave Attenuators by an Absolute Method
Author
Laverick, Elizabeth
Volume
5
Issue
4
fYear
1957
fDate
10/1/1957 12:00:00 AM
Firstpage
250
Lastpage
254
Abstract
A bridge method by means of which microwave attenuators can be calibrated absolutely is described, with a consideration of the main possible sources of error. A bridge was set up at lambda0 = 3.2 cm to test the principle of the method. It was shown that, using nonspecialized equipment, a high degree of accuracy was obtainable. An attenuator was calibrated over a range of 20 db, with an accuracy of the order of +-0.02 db. This accuracy is within the accuracy of other methods of calibration in current use, and there seems no reason why, with suitable precautions, the order of accuracy should not be improved still further, if required.
Keywords
Attenuation; Attenuators; Bridges; Calibration; Detectors; Frequency; Microwave devices; Microwave theory and techniques; Phase detection; Phase shifters;
fLanguage
English
Journal_Title
Microwave Theory and Techniques, IRE Transactions on
Publisher
ieee
ISSN
0097-2002
Type
jour
DOI
10.1109/TMTT.1957.1125160