DocumentCode :
878112
Title :
Frequency domain multiplexing for parallel acquisition of MR images
Author :
Xu, Q. ; Wang, H. ; Xu, Z. ; Li, G.
Author_Institution :
Dept. of Phys., East China Normal Univ., Shanghai, China
Volume :
42
Issue :
6
fYear :
2006
fDate :
3/16/2006 12:00:00 AM
Firstpage :
326
Lastpage :
327
Abstract :
A frequency domain multiplexing (FDM) technique is described for parallel acquisition of magnetic resonance images. Two signals from different RF coils encoded with different frequencies were combined to feed into a high-speed ADC. The multiplexed signal is then decoded by a commercial multi-carrier digital receiver. Experimentation found no phase shift, no SNR loss and essentially no crosstalk in the proposed FDM method. It is expected that the FDM technique, with the development of commercially available digital receivers, may be valuable and cost-effective in the design of a parallel MRI receiver system.
Keywords :
analogue-digital conversion; biomedical MRI; frequency division multiplexing; frequency-domain analysis; medical image processing; receivers; MR images; MRI receiver system; RF coils; frequency domain multiplexing; high-speed analog-to-digital converter; magnetic resonance images; multicarrier digital receiver; multiplexed signal; parallel acquisition;
fLanguage :
English
Journal_Title :
Electronics Letters
Publisher :
iet
ISSN :
0013-5194
Type :
jour
DOI :
10.1049/el:20063890
Filename :
1610410
Link To Document :