Title :
FMIT Diagnostic Instrumentation
Author :
Gilpatrick, J.D. ; Chamberlin, D.D.
Author_Institution :
AT-4, MS H821 Los Alamos National Laboratory, Los Alamos, NM 87545 USA
Abstract :
The Fusion Materials Irradiation Test facility (FMIT) cw prototype accelerator has noninterceptive beamline instrumentation to measure beam parameters. The transverse emittances and beam profiles are measured with an array of photodiode sensors viewing light emitted from the beam region. Tomographic reconstructions of both spatial-density distributions and of transverse-emittance distributions are performed throughout a quadrupole focusing section. Beam bunches passing through capacitive probes produce bipolar waveforms whose zero crossing corresponds to the bunch´s longitudinal centroid. By measuring the time required for a bunch to travel the known distance between two probes, velocity and energy are determined. A toroidal transformer measures the average ac beam current. Beam spill is measured by a set of movable jaws that intercept the beam edges. Each jaw contains a water flow channel whose flow rate and differential temperature are measured to derive a transverse power distribution. Beam centroid position is measured by a four-lobe, magnetic-loop pickup.
Keywords :
Instruments; Life estimation; Optical arrays; Particle beams; Photodiodes; Probes; Prototypes; Sensor arrays; Temperature measurement; Test facilities;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1985.4333781