Design and evaluation of the clear-PEM scanner for positron emission mammography
Author :
Abreu, Maria C. ; Aguiar, João D. ; Almeida, Fernando G. ; Almeida, Pedro ; Bento, Pedro ; Carriço, Bruno ; Ferreira, Miguel ; Ferreira, Nuno C. ; Gonçalves, Fernando ; Leong, Carlos ; Lopes, Filipe ; Lousã, Pedro ; Martins, Monica V. ; Matela, Nuno ; Men
Author_Institution :
LIP, Univ. do Algarve, Faro, Portugal
Volume :
53
Issue :
1
fYear :
2006
Firstpage :
71
Lastpage :
77
Abstract :
The design and evaluation of the imaging system Clear-PEM for positron emission mammography, under development by the PEM Consortium within the framework of the Crystal Clear Collaboration at CERN, is presented. The proposed apparatus is based on fast, segmented, high atomic number radiation sensors with depth-of-interaction measurement capabilities, and state-of-the-art data acquisition techniques. The camera consists of two compact and planar detector heads with dimensions 16.5×14.5 cm2 for breast and axilla imaging. Low-noise integrated electronics provide signal amplification and analog multiplexing based on a new data-driven architecture. The coincidence trigger and data acquisition architecture makes extensive use of pipeline processing structures and multi-event memories for high efficiency up to a data acquisition rate of one million events/s. Experimental validation of the detection techniques, namely the basic properties of the radiation sensors and the ability to measure the depth-of-interaction of the incoming photons, are presented. System performance in terms of detection sensitivity, count-rates and reconstructed image spatial resolution were also evaluated by means of a detailed Monte Carlo simulation and an iterative image reconstruction algorithm.
Keywords :
Monte Carlo methods; cancer; data acquisition; image reconstruction; mammography; medical image processing; pipeline processing; positron emission tomography; CERN; Monte Carlo simulation; analog multiplexing; breast cancer; data acquisition architecture; data acquisition techniques; data-driven architecture; depth-of-interaction measurement; detection sensitivity; fast segmented high atomic number radiation sensors; imaging system clear-PEM; iterative image reconstruction algorithm; low-noise integrated electronics; multievent memories; pipeline processing structures; planar detector; positron emission mammography; radiation sensors; reconstructed image spatial resolution; signal amplification; system performance; Atomic measurements; Cameras; Collaboration; Data acquisition; Detectors; Head; Image reconstruction; Image segmentation; Mammography; Radioactive decay; Avalanche photodiodes; biomedical imaging; gamma-ray detectors; image reconstruction; positron emission mammography; positron emission tomography; simulation software;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2006.870173
Filename :
1610954
Link To Document :
بازگشت