PET Detector Quality Assurance Using
Cs Singles Transmission Imaging
Author
Wassenaar, Richard W. ; DeKemp, Robert A.
Volume
53
Issue
1
fYear
2006
Firstpage
108
Lastpage
112
Abstract
Daily quality assurance of positron emission tomography detectors, to ensure accurate efficiency calibration of the system, typically requires the acquisition of coincidence data from a flood source. To obtain sufficient counts, long scan times on the order of an hour are required. The long acquisition time and necessity of an external activity source may make such a protocol impractical for daily use. A new protocol has been demonstrated for the ECAT ART scanner, using the two
Cs transmission sources. The new protocol produces a map of the counts per detector, as well as plots of the mean and standard deviation of the detector counts in a given block. Graphical outputs allow for quick recognition of problematic areas. The factory protocol requires 90 min to obtain a 1% statistical error on the individual detector counts. Due to the high activity of the
Cs sources, a 1% statistical error is obtained with an 8 min scan using the new protocol. The new protocol provides information regarding the status of all blocks, reduces acquisition time, and alleviates the need for an external activity source. The protocol presented in this study can be implemented on other
Cs transmission based scanners, and with coincidence flood sources if detector-specific singles histogramming is available.
Cs transmission sources. The new protocol produces a map of the counts per detector, as well as plots of the mean and standard deviation of the detector counts in a given block. Graphical outputs allow for quick recognition of problematic areas. The factory protocol requires 90 min to obtain a 1% statistical error on the individual detector counts. Due to the high activity of the
Cs sources, a 1% statistical error is obtained with an 8 min scan using the new protocol. The new protocol provides information regarding the status of all blocks, reduces acquisition time, and alleviates the need for an external activity source. The protocol presented in this study can be implemented on other
Cs transmission based scanners, and with coincidence flood sources if detector-specific singles histogramming is available.Keywords
137-Cs transmission; positron emission tomography (PET); quality assurance; singles transmission; Calibration; Detectors; Floods; Hospitals; Impedance; Positron emission tomography; Production facilities; Protocols; Quality assurance; Subspace constraints; 137-Cs transmission; positron emission tomography (PET); quality assurance; singles transmission;