DocumentCode :
884397
Title :
Selective Deblurring for Improved Calcification Visualization and Quantification in Carotid CT Angiography: Validation Using Micro-CT
Author :
Rollano-Hijarrubia, Empar ; Manniesing, Rashindra ; Niessen, Wiro J.
Author_Institution :
Depts. of Radiol. & Med. Inf., Erasmus MC-Univ. Med. Center of Rotterdam, Rotterdam
Volume :
28
Issue :
3
fYear :
2009
fDate :
3/1/2009 12:00:00 AM
Firstpage :
446
Lastpage :
453
Abstract :
Visualization and quantification of small structures with computed tomography (CT) is hampered by the limited spatial resolution of the system. Histogram-based selective deblurring (HiSD) is a deconvolution method that restores small high-density structures, i.e., calcifications, of a CT image, using the high-intensity voxel information of the deconvolved image, while preserving the original hounsfield Units (HUs) in the remaining tissues. In this study, high resolution micro-CT data are used to validate the potential of HiSD to improve calcium visualization and quantification in the carotid arteries on in vivo contrast-enhanced CTA data. The evaluation is performed qualitatively and quantitatively on 15 atherosclerotic plaques obtained from ten different patients. HiSD in combination with vessel segmentation significantly improves calcification visualization and quantification on in vivo contrast-enhanced CT images. Calcification blur is reduced, while avoiding noise amplification and edge-ringing artifacts in the surrounding tissues. Calcification quantification errors are reduced by 23.5% on average.
Keywords :
blood vessels; computerised tomography; diagnostic radiography; image enhancement; image segmentation; medical image processing; atherosclerotic plaques; calcification quantification; calcification visualization; carotid CT angiography; carotid arteries; computed tomography; contrast-enhanced CTA; edge-ringing artifacts; histogram-based selective deblurring; hounsfield units; image segmentation; micro-CT; noise amplification; Angiography; Calcium; Carotid arteries; Computed tomography; Data visualization; Deconvolution; Image restoration; Image segmentation; Performance evaluation; Spatial resolution; Calcification quantification; computed tomography (CT); deblurring; micro-CT validation; segmentation; Angiography; Atherosclerosis; Carotid Arteries; Carotid Stenosis; Data Interpretation, Statistical; Humans; Image Processing, Computer-Assisted; Normal Distribution; Reproducibility of Results; X-Ray Microtomography;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/TMI.2008.2006529
Filename :
4639515
Link To Document :
بازگشت