DocumentCode
887487
Title
The Performance of a Multistep Proportional Counter for Use in X-Ray Astronomy
Author
Ramsey, B.D. ; Weisskopf, M.C.
Author_Institution
Space Science Laboratory, Code ES65 NASA Marshall Space Flight Center Huntsville, Alabama 35812
Volume
34
Issue
3
fYear
1987
fDate
6/1/1987 12:00:00 AM
Firstpage
672
Lastpage
675
Abstract
The addition of a parallel field preamplification region has been shown to dramatically improve the performance of a standard multiwire proportional counter, permitting high overall gas gains to be achieved, a prerequisite for good spatial resolution, without in any way compromising energy resolution. The resulting device, known as a multistep detector, has been tested over a wide range of operating conditions and fill gas mixtures. Results are presented here detailing the performance available with argon and xenon as the principle gases, these being of particular relevance to x-ray astronomy.
Keywords
Argon; Astronomy; Counting circuits; Detectors; Energy resolution; Gases; Performance gain; Spatial resolution; Testing; Xenon;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1987.4334700