• DocumentCode
    889626
  • Title

    A room temperature setup to compare the quantized Hall resistance with 1-Ω standards

  • Author

    Boella, Giorgio ; Reedtz, G. Marullo

  • Author_Institution
    Istituto Elettrotechnico Nazionale Galileo Ferraris, Torino, Italy
  • Volume
    41
  • Issue
    1
  • fYear
    1992
  • fDate
    2/1/1992 12:00:00 AM
  • Firstpage
    59
  • Lastpage
    63
  • Abstract
    A method has been developed to relate the mean value of the Istituto Elettrotechnico Nazionale Galileo Ferraris (IEN) primary group of 1-Ω standard resistors to the i=2 quantum level of the Hall resistance. It is based on an automated potentiometric system, a 36×360-Ω Hamon network, and a commercial Kusters current comparator bridge. Descriptions of the equipment and of the measurement technique are given, together with a detailed uncertainty budget. The typical fractional total uncertainty of the measurement at the 1-σ level is 5×10-8
  • Keywords
    computerised instrumentation; electric resistance measurement; measurement standards; quantum Hall effect; 1 ohm; 20 degC; 27 degC; Hamon network; Istituto Elettrotechnico Nazionale Galileo Ferraris; Kusters current comparator bridge; automated potentiometric system; fractional total uncertainty; mean value; quantized Hall resistance; room temperature; standard resistors; uncertainty budget; Bridges; Detectors; Electrical resistance measurement; Electrons; Helium; Magnetic field measurement; Measurement uncertainty; Resistors; Standards development; Temperature;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.126632
  • Filename