• DocumentCode
    891336
  • Title

    Charge injection limits of activated iridium oxide electrodes with 0.2 ms pulses in bicarbonate buffered saline (neurological stimulation application)

  • Author

    Beebe, Xenia ; Rose, T.L.

  • Author_Institution
    EIC Labs. Inc., Norwood, MA, USA
  • Volume
    35
  • Issue
    6
  • fYear
    1988
  • fDate
    6/1/1988 12:00:00 AM
  • Firstpage
    494
  • Lastpage
    495
  • Abstract
    The maximum charge injection without electrode solution decomposition for activated iridium wire electrodes in bicarbonate buffered saline was 2.1 and 1.0 mC/cm2 geometric for anodic-first and cathodic-first, respectively, 0.2-ms balanced-charge biphasic current pulses. Electrodes biased at +0.8 V vs. SCE (saturated calomel electrode) accepted charge up to 3.5 mC/cm2 geometric with monophasic cathodal pulses.
  • Keywords
    biomedical equipment; electrodes; iridium compounds; balanced-charge biphasic current pulses; bicarbonate buffered saline; monophasic cathodal pulses; Area measurement; Counting circuits; Electrodes; Hydrogen; In vivo; Microelectrodes; Proteins; Surface cleaning; Testing; Wire; Bicarbonates; Buffers; Electric Conductivity; Electrodes; Iridium;