• DocumentCode
    892350
  • Title

    The Australian National Standards Laboratory X-Band Radiometer for the Calibration of Noise Sources

  • Author

    Somlo, P.I. ; Hollway, D.L.

  • Volume
    16
  • Issue
    9
  • fYear
    1968
  • fDate
    9/1/1968 12:00:00 AM
  • Firstpage
    664
  • Lastpage
    669
  • Abstract
    The chief possible causes of error in the calibration of microwave noise sources are examined and a radiometer designed to minimize these errors is described. This includes a self-contained hot-load standard, facilities for measuring the hot-guide losses, a synchronously switched broad-band circulator, a switchable, high-stability attenuator, provision for measuring this attenuation, and a carbon-filament lamp reference noise source.
  • Keywords
    Attenuation measurement; Attenuators; Australia; Calibration; Laboratories; Lamps; Loss measurement; Measurement standards; Microwave radiometry; Noise measurement;
  • fLanguage
    English
  • Journal_Title
    Microwave Theory and Techniques, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9480
  • Type

    jour

  • DOI
    10.1109/TMTT.1968.1126769
  • Filename
    1126769