DocumentCode :
896002
Title :
Further Analysis of the Off-Null Versus Power Ratio Method of Attenuation Measurement (Correspondence)
Author :
Larson, Winston ; Desch, R.F.
Volume :
18
Issue :
2
fYear :
1970
fDate :
2/1/1970 12:00:00 AM
Firstpage :
112
Lastpage :
113
Abstract :
Analysis of calibration data of a rotary-vane attenuator has yielded closer agreement between the recently developed off-null and the proven power ratio methods of attenuation measurement. A constant bias was discovered in the measured valures of attenuation difference, and a procedure is described to correct for this bias. After the correction is applied, the average agreement between the two methods of measurements is improved an order of magnitude, namely, from a former 2.0 percent to 0.17 percent.
Keywords :
Attenuation measurement; Attenuators; Blades; Calibration; Power measurement; Terrorism;
fLanguage :
English
Journal_Title :
Microwave Theory and Techniques, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9480
Type :
jour
DOI :
10.1109/TMTT.1970.1127163
Filename :
1127163
Link To Document :
بازگشت