DocumentCode :
896980
Title :
A Sealed High Pressure Xenon-Carbon Dioxide Filled Position Sensitive Proportional Counter
Author :
Thomas, H.D. ; Sims, M.R. ; Turner, M.J.L.
Author_Institution :
X-Ray Astronomy Group, Department of Physics, University of Leicester, Leicester, U. K.
Volume :
29
Issue :
1
fYear :
1982
Firstpage :
148
Lastpage :
150
Abstract :
A sealed position sensitive proportional counter, used in previously reported work with xenon-methane and intended for applications in X-ray astronomy, has been operated with xenon-carbon dioxide at 2 atmospheres pressure. As expected, the improved UV absorption of carbon dioxide limits the effect of secondary avalanches so that a position resolution of 230 microns FWHM is obtained compared with 360 microns in xenon-methane. The modulation of the position signal at the anode wire pitch, observed in xenon-methane is greater in xenon-carbon dioxide for the same amplifier shaping time constant. However, it does diminish for longer time constants and the position resolution improves, contrary to its behaviour in xenon-methane. Thus considerably improved position resolution with marginally degraded linearity are achieved with xenon-carbon dioxide.
Keywords :
Anodes; Astronomy; Atmosphere; Carbon dioxide; Counting circuits; Degradation; Electromagnetic wave absorption; Linearity; Signal resolution; Wire;