DocumentCode :
897252
Title :
Small-Angle X-Ray Scattering System with Linear Position-Sensitive Detector
Author :
Forouhi, A.R. ; Sleaford, B. ; Perez-Mendez, V. ; de Fontaine, D. ; Fodor, J.
Author_Institution :
Department of Materials Science and Mineral Engineering University of California, Berkeley, CA 94720
Volume :
29
Issue :
1
fYear :
1982
Firstpage :
275
Lastpage :
278
Abstract :
A linear position-sensitive detector has been constructed for use in small-angle x-ray scattering experiments. It is a one-dimensional, pressurized gasfilled detector with a delay line read-out. Calculated efficiency of the detector at 8.04 keV and 45 psi for Ar-CH4 filling is 60%, and for Xe-CO2 filling is 90% limited only by the thickness of the entrance beryllium window. Inherent resolution of the detector is 0.1 mm for normal incident x-rays. The overall resolution of the detection system is determined by the width and angular divergencies of the point collimated Cuk¿, beam used in the experiment. The performance of the system is described in terms of measurements on Al-Zn alloy samples.
Keywords :
Collimators; Delay lines; Materials science and technology; Position sensitive particle detectors; Structural beams; Synthetic aperture sonar; X-ray detection; X-ray detectors; X-ray diffraction; X-ray scattering;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1982.4335844
Filename :
4335844
Link To Document :
بازگشت