• DocumentCode
    897677
  • Title

    The Development of Mercuric Iodide Gamma-Radiation Detectors for Application in Nuclear Medicine

  • Author

    Levi, A. ; Roth, M. ; Schieber, M. ; Lavy, S. ; Cooper, G.

  • Author_Institution
    School of Applied Science and Technology the Hebrew University of Jerusalem Jerusalem, Israel
  • Volume
    29
  • Issue
    1
  • fYear
    1982
  • Firstpage
    457
  • Lastpage
    460
  • Abstract
    The feasibility of NaI(Tl) scintillators replacement by high energy resolution HgI2 detectors for rCBF measurements based on 133Xe inhalation technique has been investigated. Exhaustive phantom mapping has been carried out using 133Xe quasi-point sources. Comparative clinical rCBF measurements have been performed using the HgI2 and NaI(Tl) detectors simulaneously. The results show that the superior energy resolution of HgI2 compared to the NaI(Tl) and commercially available CdTe detectors allow efficient energy rejection of the scattered radiation, which improves considerably the spatial resolution of the rCBF instrumentation.
  • Keywords
    Energy measurement; Energy resolution; Gamma ray detection; Gamma ray detectors; Imaging phantoms; Nuclear medicine; Performance evaluation; Radiation detectors; Scattering; Spatial resolution;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1982.4335886
  • Filename
    4335886