• DocumentCode
    897824
  • Title

    Real-Time Image Reconstruction for Time-of-Flight Positron Emission Tomography (TOFPET)

  • Author

    Philippe, E.A. ; Mullani, N. ; Wong, W. ; Hartz, R.

  • Author_Institution
    University of Texas Medical School at Houston, Division of Cardiology, 6431 Fannin, Houston, TX 77030
  • Volume
    29
  • Issue
    1
  • fYear
    1982
  • Firstpage
    524
  • Lastpage
    528
  • Abstract
    Several researchers have proven the benefits of using time-of-flight data in Positron Emission Tomography (TOFPET). One of the characteristics of TOFPET which has not been previously explored is the ability to carry out real-time image reconstruction. The advantages of real-time reconstruction are four-fold: 1) Immediate visual feedback of image build-up, 2) Substantial savings of image memory, 3) Faster thoughput of patients through the system and 4) Region-of-interest reconstruction for further memory savings. We present a generalized mathematical model describing this technique together with computer simulations demonstrating its feasibility. We also discuss the implementation of this real-time reconstruction scheme in the TOFPET scanner currently under construction at our institution.
  • Keywords
    Biomedical imaging; Computer simulation; Filtering; Filters; Image reconstruction; Mathematical model; Positron emission tomography; Real time systems; Signal to noise ratio; Smoothing methods;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1982.4335900
  • Filename
    4335900