DocumentCode
897824
Title
Real-Time Image Reconstruction for Time-of-Flight Positron Emission Tomography (TOFPET)
Author
Philippe, E.A. ; Mullani, N. ; Wong, W. ; Hartz, R.
Author_Institution
University of Texas Medical School at Houston, Division of Cardiology, 6431 Fannin, Houston, TX 77030
Volume
29
Issue
1
fYear
1982
Firstpage
524
Lastpage
528
Abstract
Several researchers have proven the benefits of using time-of-flight data in Positron Emission Tomography (TOFPET). One of the characteristics of TOFPET which has not been previously explored is the ability to carry out real-time image reconstruction. The advantages of real-time reconstruction are four-fold: 1) Immediate visual feedback of image build-up, 2) Substantial savings of image memory, 3) Faster thoughput of patients through the system and 4) Region-of-interest reconstruction for further memory savings. We present a generalized mathematical model describing this technique together with computer simulations demonstrating its feasibility. We also discuss the implementation of this real-time reconstruction scheme in the TOFPET scanner currently under construction at our institution.
Keywords
Biomedical imaging; Computer simulation; Filtering; Filters; Image reconstruction; Mathematical model; Positron emission tomography; Real time systems; Signal to noise ratio; Smoothing methods;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1982.4335900
Filename
4335900
Link To Document