• DocumentCode
    898033
  • Title

    Automated Regulation of Critical Parameters and Related Design Aspects of Spectroscopy Amplifiers with Time-Invariant Filters

  • Author

    Cova, S. ; Colombo, S. ; Freitas, I.

  • Author_Institution
    Politecnico di Milano, Istituto di Fisica, 20133 Milano, Italy
  • Volume
    29
  • Issue
    1
  • fYear
    1982
  • Firstpage
    609
  • Lastpage
    613
  • Abstract
    The actual performance of modern spectroscopy amplifiers on the field is critically dependent on the setting of some analog controls and sensitive to some second-order effects, to be taken into account by the designer. The semi-automated pole-zero cancellation control is here introduced; precision adjustment is easily obtained under operating conditions by means of a Pole-Zero Adviser® circuit, without using auxiliary equipment (oscilloscope, etc.). Automated controls for the threshold and time constant of a gated baseline restorer are also considered. Some aspects in the design and high-resolution testing of spectroscopy amplifiers are discussed.
  • Keywords
    Active filters; Automatic control; Circuit testing; Degradation; Oscilloscopes; Preamplifiers; Signal processing; Signal resolution; Spectroscopy; Tail;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1982.4335920
  • Filename
    4335920