Title :
Transistor Reset Preamplifier for High Rate High Resolution Spectroscopy
Author :
Landis, D.A. ; Cork, C.P. ; Madden, N.W. ; Goulding, F.S.
Author_Institution :
Lawrence Berkeley Laboratory University of California Berkeley, California 94720 USA
Abstract :
Pulsed transistor reset of high resolution charge sensitive preamplifiers used in cooled semiconductor spectrometers can sometimes have an advantage over pulsed light reset systems. Several versions of transistor reset spectrometers using both silicon and germanium detectors have been built. This paper discusses the advantages of the transistor reset system and illustrates several configurations of the packages used for the FET and reset transistor. It also describes the preamplifer circuit and shows the performance of the spectrometer at high rates.
Keywords :
Capacitance; Crosstalk; Detectors; Energy resolution; FETs; Face detection; Feedback; Germanium; Preamplifiers; Spectroscopy;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :