Title :
A Fast Multi-Channel Analyzer for Fusion X-Ray Spectrometry
Author :
Kumahara, T. ; Shimizu, K. ; Matoba, T. ; Suzuki, Y.
Author_Institution :
Japan Atomic Energy Research Institute Tokai-mura, Naka-gun, Ibaraki-ken, Japan
Abstract :
An 128 channels CAMAC MCA with 10 MHz conversion speed has been developed for measuring spatial distribution and time dependency of fusion X-ray spectra. This CAMAC module contains a peak sensing ADC with a linearizer using four 6 bits flash type ADC IC´s (SDA 5010), a double buffered histogramming memory with multiplexers using eight 1k bits ECL RAM´s (MB 7072E) and a 10 msec setting programmable period timer. Design concept, circuit configuration and actual performances of the MCA are discussed.
Keywords :
Atomic measurements; CAMAC; Circuits; Linearity; Plasma devices; Plasma measurements; Plasma temperature; Plasma x-ray sources; Spectroscopy; Time measurement;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1982.4335988