• DocumentCode
    899251
  • Title

    An optically pumped magnetic sensor with submilligamma resolution

  • Author

    Avrin, William F. ; Sager, Ronald E.

  • Author_Institution
    Quantum Magnetics Inc., San Diego, CA, USA
  • Volume
    25
  • Issue
    5
  • fYear
    1989
  • fDate
    9/1/1989 12:00:00 AM
  • Firstpage
    3408
  • Lastpage
    3410
  • Abstract
    A free-precession magnetic detector (FPMD) has been developed using the atomic spin resonance of potassium vapor. In order to maximize sensitivity and minimize motion effects, this sensor separates the observation and spin-alignment functions in the spin resonance experiment. The FPMD exhibits a noise level of 6×10-13 T with one second of data averaging. The initial experiments suggest ways of reducing this noise to approximately 10-14 T
  • Keywords
    atomic beam magnetic resonance; magnetic field measurement; nonelectric sensing devices; optical pumping; potassium; FPMD; K vapour; atomic spin resonance; free-precession magnetic detector; noise level; optically pumped magnetic sensor; sensitivity; spin-alignment functions; submilligamma resolution; Atom optics; Detectors; Magnetic resonance; Magnetic sensors; Magnetic separation; Noise level; Noise reduction; Optical noise; Optical pumping; Optical sensors;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/20.42317
  • Filename
    42317