DocumentCode
899329
Title
High resolution position-sensitive detector for thermal neutrons
Author
Kurz, R. ; Müller, K.D. ; Schelten, J. ; Szepesvary, A. ; Schäfer, W. ; Will, G.
Author_Institution
Zentrallabor fur Elektronik, KFA Julich GmbH, West Germany
Volume
35
Issue
1
fYear
1988
Firstpage
78
Lastpage
80
Abstract
A position-sensitive scintillation detector (PSD) for thermal neutrons has been constructed. Using a /sup 6/LiI scintillator and a linear arrangement of rectangular photomultipliers, a spatial resolution of 0.5 mm has been achieved. With a dedicated hardwired computer, data are processed within 1 mu s and the differential linearity has been improved. The gamma -sensitivity is drastically reduced in comparison with /sup 6/Li glass scintillator detectors.<>
Keywords
gamma-ray detection and measurement; neutron detection and measurement; position sensitive particle detectors; scintillation counters; /sup 6/LiI; 0.5 mm; 1 mus; differential linearity; gamma -sensitivity; position-sensitive scintillation detector; rectangular photomultipliers; spatial resolution; thermal neutrons; Gamma ray detection; Gamma ray detectors; Glass; Linearity; Neutrons; Photomultipliers; Position sensitive particle detectors; Scintillation counters; Solid scintillation detectors; Spatial resolution;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.12677
Filename
12677