DocumentCode :
900455
Title :
Dielectric measurements with an open-ended coaxial probe
Author :
Marsland, T.P. ; Evans, S.
Author_Institution :
Cambridge University, Engineering Department, Cambridge, UK
Volume :
134
Issue :
4
fYear :
1987
fDate :
8/1/1987 12:00:00 AM
Firstpage :
341
Lastpage :
349
Abstract :
The paper is concerned with the use of an open-ended coaxial-line probe for making dielectric measurements on lossy materials. The input reflection coefficient of the probe is measured with a network analyser, and the permittivity is determined using an approximate circuit model for the probe admittance. Applying the cross-ratio invariance property of the bilinear transform to two such circuit models, we derive a simplified procedure that permits simultaneous calibration of the probe and error correction of the network analyser. Estimates of the resulting uncertainties are given, and the technique is illustrated with experimental measurements on known dielectric liquids between 50 MHz and 2.6 GHz.
Keywords :
calibration; dielectric measurement; microwave measurement; network analysers; permittivity measurement; probes; 50 MHz to 2.6 GHz; bilinear transform; calibration; cross-ratio invariance property; dielectric liquids; dielectric measurements; error correction; input reflection coefficient; lossy materials; network analyser; open-ended coaxial-line probe; permittivity; probe admittance;
fLanguage :
English
Journal_Title :
Microwaves, Antennas and Propagation, IEE Proceedings H
Publisher :
iet
ISSN :
0950-107X
Type :
jour
DOI :
10.1049/ip-h-2.1987.0068
Filename :
4643028
Link To Document :