• DocumentCode
    901528
  • Title

    An implantable multielectrode array with on-chip signal processing

  • Author

    Najafi, Khalil ; Wise, Kensall D.

  • Volume
    21
  • Issue
    6
  • fYear
    1986
  • fDate
    12/1/1986 12:00:00 AM
  • Firstpage
    1035
  • Lastpage
    1044
  • Abstract
    This active probe can be used for the long-term recording of extracellular neural biopotentials and as a basis for closed-loop neural prostheses. The probe incorporates on-chip circuitry for amplifying, multiplexing, and buffering neural signals recorded from ten recording electrodes spaced 100-μm apart. It requires only three leads and operates from a single 5-V supply. On-chip self-test circuitry for testing electrode impedance levels is provided. The on-chip circuitry is fabricated in a die area of 1.3 mm/SUP 2/ using 6-μm LOCOS enhancement-depletion NMOS technology, and dissipates 5 mW of power. The probe is 4.7 mm long and 15 μm thick, and has a shank which tapers from 160 μm near the base to less than 15 μm near the tip.
  • Keywords
    Bioelectric potentials; Biomedical electronics; Biomedical equipment; Field effect integrated circuits; Probes; Prosthetics; bioelectric potentials; biomedical electronics; biomedical equipment; field effect integrated circuits; probes; prosthetics; Array signal processing; Automatic testing; Biomedical signal processing; Built-in self-test; Circuit testing; Electrodes; Extracellular; Impedance; Probes; Prosthetics;
  • fLanguage
    English
  • Journal_Title
    Solid-State Circuits, IEEE Journal of
  • Publisher
    ieee
  • ISSN
    0018-9200
  • Type

    jour

  • DOI
    10.1109/JSSC.1986.1052646
  • Filename
    1052646