• DocumentCode
    905526
  • Title

    Reduction of the electronics of a 1D position sensitive detector

  • Author

    Duijn, J.H. ; van Eijk, C.W.E. ; Hollander, R.W. ; Sloof, G.W.

  • Author_Institution
    Dept. of Appl. Phys., Delft Univ. of Technol., Netherlands
  • Volume
    35
  • Issue
    1
  • fYear
    1988
  • Firstpage
    135
  • Lastpage
    137
  • Abstract
    A readout method has been developed for a one-dimensional position-sensitive X-ray detector that reduces the readout electronics significantly. The method combines features of cathode-strip and cathode-wedge detector readouts. The readout electronics of the authors´ detector, which has 36 cathode strips, has been reduced from 36 to only eight independent channels without loss of spatial resolution. A diffraction spectrum of corundum recorded with the detector system is shown.<>
  • Keywords
    X-ray detection and measurement; position sensitive particle detectors; 1D position sensitive detector; Al/sub 2/O/sub 3/; X-ray detector; cathode-strip; cathode-wedge detector readouts; corundum; diffraction spectrum; electronics; readout method; Anodes; Blades; EPROM; Hardware; Linearity; Position sensitive particle detectors; Quantum computing; Signal resolution; Strips; Table lookup;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.12691
  • Filename
    12691