Title :
Investigation of Large Format Microchannel Plate Z Configurations
Author :
Siegmund, O.H.W. ; Coburn, K. ; Malina, R.F.
Author_Institution :
Space Sciences Laboratory University of California, Berkeley
Abstract :
The performance of triplet (Z) stacks of microchannel plates (MCP´s) has been studied as a part of the instrument development for the Extreme Ultraviolet Explorer [1] (EUVE) satellite mission. Relatively large MCP´s with a 60 mm diameter and having an 80:1 channel length to diameter (L/D) ratio were used in several configurations. These have been used in the EUVE prototype imaging detector to provide > 512 Ã 512 pixels with low image distortion (< 1%). We have examined the gain and pulse height characteristics showing that both high gains (> 2 Ã 107) and tight output pulse height distributions (< 30% FWHM) may be achieved. Simple distribution techniques have also allowed low intrinsic background event rates (< .15 events cm-2 sec-1) to be obtained. Variation of the quantum efficiency of MCP´s over the wavelength range 160 Ã
to 1216 Ã
has been investigated for a range of angles of incidence. The effect of temperature variations on MCP operating characteristics has also been evaluated.
Keywords :
Anodes; Clouds; Detectors; Electrons; Feedback; Laboratories; Microchannel; Space missions; Strips; Ultraviolet sources;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1985.4336871