• DocumentCode
    907020
  • Title

    Scatter Compensation Techniques for SPECT

  • Author

    Jaszczak, Ronald J. ; Floyd, Carey E., Jr. ; Coleman, R. Edward

  • Author_Institution
    Duke University Medical Center Durham, North Carolina
  • Volume
    32
  • Issue
    1
  • fYear
    1985
  • Firstpage
    786
  • Lastpage
    793
  • Abstract
    Compton scattered photons included within the photopeak pulse-height window decrease the image contrasts of lesions. Furthermore, in the absence of an effective compensation procedure, the accuracy of the SPECT measurement is reduced. In the following sections the effect of scatter on SPECT is reviewed, and the influence of scattered photons on the effective linear attenuation coefficient used with multiplicative attenuation compensation methods is described. Criteria for evaluating scatter compensation procedures are proposed, and approaches to reducing the effect of scatter on SPECT imaging are described.
  • Keywords
    Attenuation; Cameras; Degradation; Electromagnetic scattering; Electrons; Frequency; Lesions; Particle scattering; Single photon emission computed tomography; X-ray scattering;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1985.4336941
  • Filename
    4336941