• DocumentCode
    907104
  • Title

    Development of a High Resolution Positron CT for Animal Studies

  • Author

    Tomitani, T. ; Nohara, N. ; Murayama, H. ; Yamamoto, M. ; Tanaka, E.

  • Author_Institution
    Division of Physics National Institute of Radiological Sciences, 9-1, 4-Chome-Anagawa, Chiba-shi, 260, Japan
  • Volume
    32
  • Issue
    1
  • fYear
    1985
  • Firstpage
    822
  • Lastpage
    825
  • Abstract
    A single ring high resolution positron CT for animal studies was developed which utilizes 64 gridded rectangular photomultipliers (l3mm square) and 128 rectangular Bismuth Germanate crystal (4mm wide, 10mm high and 20mm long). A photomultiplier is incorporated with a grid near the photocathode which controls photoelectron transport of half part of a tube. This enalbes us the identification of the fired crystal out of two coupled to a photomultiplier. Detector ring, animal port and field of view diameters are 265, 135 and 128mm, respectively. Detector arrangement was determined by POSITOLOGY. The spatial resolutions in fwhm after unfolding source spread (2mm dia.) were 2.2mm at center and 3.3mm (tangential) and 3.6mm(radial) at 60mm from the center. Hardware out-of-FOV event rejector was implemented which rejects events outside the field of view and reduces event rate by a factor of 3 in the case of calibration measurements.
  • Keywords
    Animals; Calibration; Cathodes; Computed tomography; Crystals; Detectors; Photomultipliers; Physics; Positrons; Spatial resolution;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1985.4336948
  • Filename
    4336948