• DocumentCode
    907377
  • Title

    Computer Simulation of Semiconductor Dosimeters

  • Author

    Manfredotti, C. ; Nastasi, U. ; Zanini, A. ; Bonazzola, G.C.

  • Author_Institution
    Istituto di Fisica Superiore, UniversitÃ\xa0 di Torino and Istituto Nazionale di Fisica Nucleare, Sezione di Torino Corso M. D´´Azeglio 46, 10125 Torino (Italy)
  • Volume
    32
  • Issue
    1
  • fYear
    1985
  • Firstpage
    953
  • Lastpage
    956
  • Abstract
    A previous Monte Carlo program, ISIDE, for the simulation of semiconductor gamma ray detectors, has been extended to a new code, PHANTOM, which can be used either for calculating the gamma ray exposure from the output of semiconductor detectors working in pulse mode or for determining the isodose curves in a cylindrical phantom exposed to a disk gamma ray source. The results are used to evaluate the exposure from the spectra recorded by a HgI2 detector in the energy range 10 KeV - 1 MeV. A good agreement with experimental results either for phantoms and for semiconductor detectors is obtained.
  • Keywords
    Apertures; Circuit simulation; Collimators; Computational modeling; Computer simulation; Filters; Gamma ray detection; Gamma ray detectors; Imaging phantoms; Monte Carlo methods;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1985.4336973
  • Filename
    4336973