DocumentCode
907429
Title
New Method of Proportional Counter Feedback Biasing for Wide-Range Radiation Dose-Rate Monitors
Author
Kopp, M.K. ; Valentine, K.H. ; Guerrant, G.C. ; Manning, F.W.
Author_Institution
Instrumentation and Controls Division Oak Ridge National Laboratory Oak Ridge, Tennessee 37831, USA
Volume
32
Issue
1
fYear
1985
Firstpage
966
Lastpage
968
Abstract
A prototypic wide-range radiation dose-rate monitor for civil defense applications has been developed and tested. The specified dose-rate range (0 to 500 R/h) was displayed on a single readout scale by using feedback-controlled biasing of a proportional counter. This new method is based on controlling the avalanche multiplication factor (gas gain) of the counter by varying its bias voltage in response to its measured output current (i.e., detected dose rate). The counter output current varies between 0 and 1.5 nA in a quasilogarithmic response to dose rates between 0 and 500 R/h. The corresponding values of gas gain and bias voltage range from 1 to 300 and 200 to 1900 V respectively.
Keywords
Current measurement; Feedback; Gain measurement; Instruments; Ionization; Proportional control; Prototypes; Radiation detectors; Testing; Voltage control;