• DocumentCode
    908685
  • Title

    High Resolution X - Gamma-Ray Spectrometry at High Counting Rates Using High Purity Germanium Detectors

  • Author

    Howes, J.H. ; Allsworth, F.L.

  • Author_Institution
    Instrumentation and Applied Physics Division, AERE Harwell, OXON OX11 ORA, U.K.
  • Volume
    33
  • Issue
    1
  • fYear
    1986
  • Firstpage
    283
  • Lastpage
    287
  • Abstract
    One of the limitations in achieving high energy resolution in X and gamma-ray spectrometers at high counting rates is incomplete charge collection in the germanium detector. In this paper we discuss some of the limiting factors which are relevant in detector fabrication and in the optical reset configurations in the input integrator circuit and the associated electronic processing. Experimental results demonstrate input counting rates up to 1.5 million counts per sec without any serious degradation in energy resolution.
  • Keywords
    Circuits; Degradation; Energy resolution; Gamma ray detection; Gamma ray detectors; Geometry; Germanium; Light emitting diodes; Spectroscopy; Switches;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1986.4337100
  • Filename
    4337100