DocumentCode :
908742
Title :
Performance of a Gamma-Ray and X-Ray Spectrometer Utilizing Germanium and Si(Li) Detectors Cooled by a Closed-Cycle Cryogenic Mechanical Refrigerator
Author :
Stone, R.E. ; Barkley, V.A. ; Fleming, J.A.
Author_Institution :
EG&G ORTEC 100 Midland Road Oak Ridge, Tennessee 37830 U.S.A.
Volume :
33
Issue :
1
fYear :
1986
Firstpage :
299
Lastpage :
302
Abstract :
A mechanically cooled spectrometer has been constructed and tested separately with Si(Li) planar, high-purity germanium (HPGe) planar, and HPGe coaxial detectors. For each of these types of semiconductor detectors, this spectrometer has attained the lowest noise and best resolution at 5.9 keV, 122 keV, or 1.33 MeV of any closed-cycle mechanically cooled spectrometer which has been reported on in the literature. Resolutions at 5.9 keV were 175 eV FWHM with a 4 mm diameter Si(Li) detector and 202 eV with a 16 mm HPGe planar detector. With a 10% relative efficiency HPGe coaxial detector, resolutions of 816 eV at 122 keV and 1.70 keV at 1.33 MeV were obtained.
Keywords :
Coaxial components; Cryogenics; Gamma ray detection; Gamma ray detectors; Germanium; Refrigeration; Spectroscopy; Testing; X-ray detection; X-ray detectors;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1986.4337104
Filename :
4337104