DocumentCode
909241
Title
Design of a Mosaic BGO Detector System for Positron CT
Author
Uchida, H. ; Yamashita, T. ; Iida, M. ; Muranatsu, S.
Author_Institution
Hamamatsu Photonics K.K. 1126-1 Ichino-cho Hamamatsu-City, Japan
Volume
33
Issue
1
fYear
1986
Firstpage
464
Lastpage
467
Abstract
A new scintillation detector system has been designed for application in high spatial resolution positron CT (PCT) system. Many small BGO crystals are closely packed and coupled to a position-sensitive photomultiplier tube (PMT). The detector performance and feasibility for PCT are discussed based on both calculations and experiments. The coincidence response function of approximately 3.2mm FWHM and coincidence time resolution of less than 6ns FWHM were obtained. The results promise the possibility of a new high resolution PCT system.
Keywords
Anodes; Geometry; Light scattering; Optical arrays; Photonic crystals; Positrons; Scintillation counters; Sensor arrays; Solid scintillation detectors; Spatial resolution;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1986.4337144
Filename
4337144
Link To Document