DocumentCode :
909429
Title :
A Real Time Digital Coincidence Processor for Positron Emission Tomography
Author :
Dent, H.M. ; Jones, W.F. ; Casey, M.E.
Author_Institution :
Computer Technology and Imaging, Inc. 215 Center Park Drive Suite 1500 Knoxville, Tennessee 37922
Volume :
33
Issue :
1
fYear :
1986
Firstpage :
556
Lastpage :
559
Abstract :
A Real Time Digital Coincidence Processor has been developed for use in the Positron Emission Tomograph (PET) ECAT® scanners manufactured by Computer Technology and Imaging, Inc. (CTI). The primary functions of the Coincidence Processor include: receive from the BGO detector modules serial data, which includes timing information and detector identification; process the received data to form coincidence detector pairs; and present the coincidence pair data to a Real Time Sorter . The primary design emphasis was placed on the Coincidence Processor being able to process the detector data into coincidence pairs at real time rates. This paper will briefly describe the Coincidence Processor and some of the considerations that went into its design.
Keywords :
Cables; Clocks; Computer aided manufacturing; Counting circuits; Detectors; Positron emission tomography; Sensor arrays; Shift registers; Signal processing; Timing;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1986.4337164
Filename :
4337164