DocumentCode :
910919
Title :
Feedback Threshold Control Extends Pulse Rate Measurements into the Pileup Region
Author :
Valentine, K.H. ; Kopp, M.K. ; Williams, J.A.
Author_Institution :
Instrumentation and Controls Division Oak Ridge National Laboratory, Bldg. 3500, Oak Ridge, TN 37831-6008 USA
Volume :
34
Issue :
1
fYear :
1987
Firstpage :
77
Lastpage :
81
Abstract :
By using feedback control of the discrimination threshold we have added 4-6 decades to the useful range of otherwise conventional pulse counting systems. The additional range is obtained at the high end by retarding saturation of the output count rate with higher threshold levels. We have implemented the method with low-power microcircuits that provide a 9-decade range (10-5 to 104 R/h) without the need for range or mode switching.
Keywords :
Detectors; Diodes; Event detection; Feedback; Instruments; Laboratories; Nuclear and plasma sciences; Pulse amplifiers; Pulse measurements; Silicon;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1987.4337305