• DocumentCode
    911529
  • Title

    Single Photon Emission Computed Tomography with Increased Sampling Density at Central Region of Field-of-View

  • Author

    Nohara, Norimasa ; Murayama, Hideo ; Tanaka, Eiichi

  • Author_Institution
    Division of Physics National Institute of Radiological Sciences 9-1, Anagawa-4-chome, Chiba-shi 260, Japan
  • Volume
    34
  • Issue
    1
  • fYear
    1987
  • Firstpage
    359
  • Lastpage
    363
  • Abstract
    Single photon emission computed tomography (SPECT) systems are proposed which provide images with improved signal-to-noise ratio. The systems consist of gamma cameras equipped with collimators which view an object to be imaged with an increased sampling density in the vicinity of the center of a field-of-view. The detection systems produce weighted projection data. Evaluation was made with respect to noise characteristics in images resulting from such systems. Simulation studies showed that dual and quadruple camera systems equipped with a combination of fan beam collimators result in noise reduction of about 37% at the center of image and uniform noise over an image area, compared with systems equipped with parallel hole collimators. This suggests that the proposed systems have a possibility of improvement on quantitation of SPECT.
  • Keywords
    Attenuation; Cameras; Collimators; Image quality; Image reconstruction; Image sampling; Noise reduction; Sampling methods; Signal to noise ratio; Single photon emission computed tomography;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1987.4337363
  • Filename