• DocumentCode
    915087
  • Title

    Method of power recycling in CoAxial Mach-Zehnder interferometers for low noise measurements

  • Author

    Parker, Stephen ; Ivanov, Eugene ; Tobar, Michael

  • Author_Institution
    Sch. of Phys., Univ. of Western Australia, Crawley, WA
  • Volume
    56
  • Issue
    5
  • fYear
    2009
  • fDate
    5/1/2009 12:00:00 AM
  • Firstpage
    921
  • Lastpage
    925
  • Abstract
    We present the first experimental study of a new type of power recycling microwave interferometer designed for low noise measurements. This system enhances sensitivity to phase fluctuations in a device under test, independent of input power levels. The single sideband thermal white phase noise floor of the system has been lowered by 8 dB (reaching -185 dBc/Hz at 1 kHz offset frequency) at relatively low power levels (13 dBm).
  • Keywords
    Mach-Zehnder interferometers; electric noise measurement; fluctuations; coaxial Mach-Zehnder interferometers; device under test; low noise measurements; phase fluctuations; power recycling microwave interferometer; thermal white phase noise floor; Attenuation; Coaxial components; Cryogenics; Fluctuations; Insertion loss; Noise measurement; Phase noise; Recycling; Temperature; Voltage;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/TUFFC.2009.1124
  • Filename