    Quantitative ultrasound backscatter for pulsed cavitational ultrasound therapy-histotripsy

  • Author

    Wang, Tzu-Yin ; Xu, Zhen ; Winterroth, Frank ; Hall, Timothy L. ; Fowlkes, J. Brian ; Rothman, Edward D. ; Roberts, William W. ; Cain, Charles A.

  • Author_Institution
    Dept. of Biomed. Eng., Univ. of Michigan, Ann Arbor, MI
  • Volume
    56
  • Issue
    5
  • fYear
    2009
  • fDate
    5/1/2009 12:00:00 AM
  • Firstpage
    995
  • Lastpage
    1005
  • Abstract
    Histotripsy is a well-controlled ultrasonic tissue ablation technology that mechanically and progressively fractionates tissue structures using cavitation. The fractionated tissue volume can be monitored with ultrasound imaging because a significant ultrasound backscatter reduction occurs. This paper correlates the ultrasound backscatter reduction with the degree of tissue fractionation characterized by the percentage of remaining normal-appearing cell nuclei on histology. Different degrees of tissue fractionation were generated in vitro in freshly excised porcine kidneys by varying the number of therapeutic ultrasound pulses from 100 to 2000 pulses per treatment location. All ultrasound pulses were 15 cycles at 1 MHz delivered at 100 Hz pulse repetition frequency and 19 MPa peak negative pressure. The results showed that the normalized backscatter intensity decreased exponentially with increasing number of pulses. Correspondingly, the percentage of normal appearing nuclei in the treated area decreased exponentially as well. A linear correlation existed between the normalized backscatter intensity and the percentage of normal appearing cell nuclei in the treated region. This suggests that the normalized backscatter intensity may be a potential quantitative real-time feedback parameter for histotripsy-induced tissue fractionation. This quantitative feedback may allow the prediction of local clinical outcomes, i.e., when a tissue volume has been sufficiently treated.
  • Keywords
    backscatter; biological tissues; biomedical ultrasonics; correlation methods; kidney; surgery; ultrasonic imaging; ultrasonic therapy; cavitational ultrasound therapy; cell nuclei; frequency 1 MHz; frequency 100 Hz; histotripsy; kidney; linear correlation; tissue structure fractionation; ultrasonic tissue ablation technology; ultrasound backscatter; ultrasound imaging; Acoustic pulses; Acoustic transducers; Backscatter; Biomedical engineering; Cells (biology); Circuits; Focusing; Fractionation; Ultrasonic imaging; Ultrasonic transducers;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/TUFFC.2009.1131
  • Filename
    4976284