• DocumentCode
    916771
  • Title

    Investigation of high purity Si detectors with ohmic contacts at low temperature for dark matter searches

  • Author

    Spooner, N.J.C. ; Homer, G.J. ; Smith, P.F. ; Bewick, A.

  • Author_Institution
    Astrophys. & Nucl. Phys. Lab., Oxford Univ., UK
  • Volume
    40
  • Issue
    3
  • fYear
    1993
  • fDate
    6/1/1993 12:00:00 AM
  • Firstpage
    275
  • Lastpage
    279
  • Abstract
    The development of dark matter ionization/thermal hybrid detectors using silicon requires substrate material in which good charge collection can be achieved at low applied voltages (typically <3 V) and low temperature (typically <300 mK). Presented are results from studies of the effect of material purity on the charge collection efficiency and sensitive volume of Si ionization detectors, fabricated with low resistance contacts and operated at low temperature and bias. Samples of float-zone hypercube Si with four-point room temperature resistivities of 6, 15, 90, and 140 kΩ-cm were used. At applied voltages below 3 V improved detector characteristics were observed with the higher resistivity material. In particular, at 1 V bias a detector of 0.5-mm thickness using 140-kΩ-cm material achieved a factor of two times greater charge collection than a similar detector using 6-Ω-cm material
  • Keywords
    ionisation chambers; ohmic contacts; semiconductor counters; 0.5 mm; 3 V; 300 mK; Si ionization detectors; charge collection efficiency; dark matter; float-zone hypercube Si; high purity Si detectors; low resistance contacts; low temperature; ohmic contacts; sensitive volume; Conductivity; Cosmic rays; Detectors; Electrons; Ionization; Ocean temperature; Ohmic contacts; Pollution measurement; Silicon; Temperature sensors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.221051
  • Filename