DocumentCode :
916810
Title :
SPECT reconstruction of combined cone beam and parallel hole collimation with experimental data
Author :
Li, Jianying ; Jaszczak, Ronald J. ; Turkington, Timothy G. ; Greer, Kim L. ; Coleman, R. Edward
Author_Institution :
Dept. of Radiol., Duke Univ. Med, Center, Durham, NC, USA
Volume :
40
Issue :
3
fYear :
1993
fDate :
6/1/1993 12:00:00 AM
Firstpage :
300
Lastpage :
306
Abstract :
Three methods for combining parallel and cone beam (P&CB) SPECT data using modified maximum-likelihood expectation-maximization (ML-EM) algorithms are presented. The first method applies both parallel and cone beam data sets to reconstruct a single intermediate image after each iteration using the ML-EM algorithm. The other two are iterative methods that combine the intermediate parallel beam and cone beam source estimates to enhance the uniformity of images. These two methods are ad hoc methods. These combined collimation methods are qualitatively evaluated using experimental data. Attenuation compensation is performed by including the effects of attenuation in the transition matrix as a multiplicative factor. The results indicate that the combined P&CB approaches suppress artifacts caused by truncated projections and corrects for th distortions of the CB-only images
Keywords :
computerised tomography; image reconstruction; maximum likelihood estimation; medical signal processing; radioisotope scanning and imaging; ML-EM algorithm; attenuation; combined cone beam and parallel hole collimation; cone beam SPECT; maximum likelihood expectation maximization algorithm; parallel SPECT; transition matrix; truncated projections; Attenuation; Cameras; Cancer; Geometry; Image reconstruction; Iterative algorithms; Iterative methods; Lesions; Optical collimators; Single photon emission computed tomography;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.221055