DocumentCode :
917310
Title :
Performance Characteristics of an Automated Broad-Band Bolometer Unit Calibration System
Author :
Komarek, Ernest L.
Volume :
25
Issue :
12
fYear :
1977
fDate :
12/1/1977 12:00:00 AM
Firstpage :
1122
Lastpage :
1127
Abstract :
The arbitrary six-port reflectometer concept has been applied to an automated broad-band system in the 1-18-GHz frequency range for the calibration of bolometer units. Performance evaluation results show an improvement in precision over other automated and manually operated measurement systems used at the National Bureau of Standards. Initial evaluation results reported here show a single measurement standard deviation of 0.02 to 0.41 percent from 2-18 GHz. The system is currently used for the calibration of coaxial and waveguide bolometer unit effective efficiency.
Keywords :
Bolometers; Calibration; Coaxial components; Frequency measurement; Microwave measurements; Microwave theory and techniques; NIST; Power measurement; RF signals; Testing;