• DocumentCode
    920155
  • Title

    Improvement in open-resonator permittivity measurement

  • Author

    Cullen, A.L. ; Nagenthiram, P. ; Williams, A.D.

  • Author_Institution
    University College London, Department of Electronic & Electrical Engineering, London, UK
  • Volume
    8
  • Issue
    23
  • fYear
    1972
  • Firstpage
    577
  • Lastpage
    579
  • Abstract
    The accuracy of permittivity measurement using an open resonator has been studied. The error in the determination of permittivity arising from the imperfection of beam-wave theory is estimated. It is further shown that, by an improved method of analysing the experimental results, the effect of the imperfection in the theory is reduced and a great improvement in accuracy can be obtained.
  • Keywords
    measurement errors; permittivity measurement; resonators; beam wave theory; measurement errors; open resonators; permittivity measurements;
  • fLanguage
    English
  • Journal_Title
    Electronics Letters
  • Publisher
    iet
  • ISSN
    0013-5194
  • Type

    jour

  • DOI
    10.1049/el:19720419
  • Filename