DocumentCode :
922414
Title :
Performance of a Dual Six-Port Automatic Network Analyzer
Author :
Hoer, Cletus A.
Volume :
27
Issue :
12
fYear :
1979
fDate :
12/1/1979 12:00:00 AM
Firstpage :
993
Lastpage :
998
Abstract :
lnitial results of the performance of an experimental dual six-port automatic network analyzer operating in the 2-18-GHz range with thermistor-type power detectors are given. The imprecision in measuring reflection coefficients of one-port devices, or the scattering parameters of two-port devices is 4x10-5, excluding connector repeatability. At 3 GHz, the imprecision in measuring attenuation varies from 0.0003 dB at low values of attenuation to 0.15 dB at 60 dB. The systematic error in measuring attenuation appears to be Iess than the imprecision. The systematic error in measuring reflection coefficient appears to be less than 0.0004. Additional systematic errors caused by changes in the calibration constants over a 20-week period were observed to be less than 0.003 dB in attenuation and less than 0.002 in reflection coefficient.
Keywords :
Attenuation measurement; Calibration; Connectors; Detectors; NIST; Performance analysis; Phase shifters; Reflection; Stability; Transmission line measurements;
fLanguage :
English
Journal_Title :
Microwave Theory and Techniques, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9480
Type :
jour
DOI :
10.1109/TMTT.1979.1129779
Filename :
1129779