DocumentCode :
922884
Title :
Use of strain gauge for measuring voltage
Author :
Singh, V.R. ; Parshad, R. ; Gupta, U.C.
Volume :
60
Issue :
6
fYear :
1972
fDate :
6/1/1972 12:00:00 AM
Firstpage :
740
Lastpage :
740
Abstract :
The force of attraction between the capacitor plates is used to measure voltage by making one of the plates deflectable under the force of attraction. A strain gauge mounted on this plate suffers a shear strain which is detected electrically by the conventional technique of the Wheatstone bridge followed by an amplifier. The output of the amplifier was found to be linearly related to the square of applied voltage between the plates, as it theoretically should be. The technique can be used to measure dc and ac voltages, particularly when their magnitudes are high and there is also high impedance.
Keywords :
Bridge circuits; Capacitive sensors; Capacitors; Electrons; Force measurement; Impedance measurement; Mechanical variables measurement; Particle measurements; Strain measurement; Voltage measurement;
fLanguage :
English
Journal_Title :
Proceedings of the IEEE
Publisher :
ieee
ISSN :
0018-9219
Type :
jour
DOI :
10.1109/PROC.1972.8748
Filename :
1450678