Title :
Diathermy Applicators with Circular Aperture and Corrugated Flange
Author :
Stuchly, Maria A. ; Stuchly, Stanislaw S. ; Kantor, Gideon
fDate :
3/1/1980 12:00:00 AM
Abstract :
A design method and experimental results for a direct-contact circular aperture applicator are provided. The aperture is excited in the TE11 mode; a corrugated flange surrounding the aperture improves the uniformity of the heating pattern and limits leakage. The performance of the applicators operating in the S band (2.45 GHz) and the X band (9.96 GHz) has been tested using a short monopole probe and a thermographic camera. The heating patterns obtained by the two methods are in agreement within the experimental errors. The applicators are suitable for clinical use, as they are lightweight and rugged, and capable of delivering a desired energy dose effectively, thanks to a relatively small standing-wave ratio (SWR < 2) and very low leakage.
Keywords :
Apertures; Applicators; Couplings; Curve fitting; Electromagnetic heating; Flanges; Frequency; Microwave devices; Muscles; Skin neoplasms;
Journal_Title :
Microwave Theory and Techniques, IEEE Transactions on
DOI :
10.1109/TMTT.1980.1130054