• DocumentCode
    924289
  • Title

    High-precision measurement of 2e/hthat uses cryogenic electronic components

  • Author

    Gallop, J.C. ; Petley, B.W.

  • Author_Institution
    National Physical Laboratory, Division of Quantum Metrology, Teddington, UK
  • Volume
    9
  • Issue
    21
  • fYear
    1973
  • Firstpage
    488
  • Lastpage
    489
  • Abstract
    A new determination of 2e/h is described, in which not only the Josephson junction but also the current-null detector and 100:1 ratio voltage divider are operated at 4.2 K. The results obtained with this new system are compared with previous data obtained with room-temperature components, and the agreement is found to be 0.02 (±0.05) parts in 106.
  • Keywords
    Josephson junctions; constants; cryogenics; measurement systems; Josephson junction; constants; cryogenics; current null detector; high precision measurement; voltage divider;
  • fLanguage
    English
  • Journal_Title
    Electronics Letters
  • Publisher
    iet
  • ISSN
    0013-5194
  • Type

    jour

  • DOI
    10.1049/el:19730359
  • Filename
    4236312