• DocumentCode
    926928
  • Title

    Delay line readout of microchannel plates

  • Author

    Sobottka, Stanley E. ; Williams, Mark B.

  • Author_Institution
    Dept. of Phys., Virginia Univ., Charlottesville, VA, USA
  • Volume
    35
  • Issue
    1
  • fYear
    1988
  • Firstpage
    348
  • Lastpage
    351
  • Abstract
    A two-dimensional delay line readout to be used with large-diameter (127-mm) microchannel plates (MCPs) has been developed. High spatial resolutions (25 mu m FWHM for an MCP gain of 2*10/sup 7/) and excellent linearity of position readout (less than +or-65- mu m deviation from linearity) have been achieved in a test setup using 30-mm-diameter MCPs coupled to 140-mm*140-mm delay lines. Single-pass delays of only 176 ns allow system deadtimes of less than 1 mu s. This type of readout will be applied to imaging X-ray detectors to be used in protein crystallography.<>
  • Keywords
    delay lines; electron multipliers; image intensifiers; phototubes; 1 mus; 176 ns; 25 mum; 30 mm; deadtimes; delay line readout; imaging X-ray detectors; microchannel plates; position readout; protein crystallography; spatial resolutions; Delay lines; Insulation; Linearity; Microchannel; Optical imaging; Proteins; Spatial resolution; Wounds; X-ray detectors; X-ray imaging;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.12740
  • Filename
    12740