Abstract :
Temperature measurements in fat-muscle phantoms using thermography and a minimally perturbing temperature probe were investigated. Two microwave applicators (915- and 2450-MHz) were used to induce the heating in the phantom. Discrepancies between data taken with the thermographic camera versus the probe were measured. These discrepancies were shown to be primarily caused by a 40-s time delay in performing temperature measurements with the thermographic camera, which resulted in additional thermal diffusion in the phantom.
Keywords :
Antenna measurements; Applicators; Cameras; Electromagnetic heating; Imaging phantoms; Microwave measurements; Power measurement; Probes; Temperature measurement; Thermal conductivity;