A High-Power Dual Six-Port Automatic Network Analyzer Used in Determining Biological Effects of Rf and Microwave Radiation
Author
Hoer, Cletus A.
Volume
29
Issue
12
fYear
1981
fDate
12/1/1981 12:00:00 AM
Firstpage
1356
Lastpage
1364
Abstract
The design, calibration, and performance of a high-power (1-1000 W) automatic network analyzer based on the six-port concept are described for the 10-100-MHz range. Calibration is performed with a length of transmission line as the only impedence standard needed. A 10-mW thermistor mount is the standard of power. Imprecision in measnring reflection coefficient Gamma is 0.0001 in magnitude and 0.005/|Gamma| degrees in phase. Corresponding estimated systematic errors are 0.001 and 0.1/|Gamma| degrees. Imprecision in measuring power is 0.01 percent of range (20 W, 200 W, or 1000 W) with an estimated systematic error of 1.25 percent of reading.
Keywords
Calibration; Detectors; Magnetic field measurement; Performance analysis; Phase detection; Power measurement; Radio frequency; Reflection; Synthesizers; Testing;