• DocumentCode
    931973
  • Title

    Strategies for Generating Prolonged Functional Standing Using Intramuscular Stimulation or Intraspinal Microstimulation

  • Author

    Lau, Bernice ; Guevremont, Lisa ; Mushahwar, Vivian K.

  • Author_Institution
    Univ. of Alberta, Edmonton
  • Volume
    15
  • Issue
    2
  • fYear
    2007
  • fDate
    6/1/2007 12:00:00 AM
  • Firstpage
    273
  • Lastpage
    285
  • Abstract
    Spinal cord injury (SCI) often results in the loss of the ability to stand. The goal of this study was to implement a functional electrical stimulation (FES) system for restoring prolonged periods of standing after SCI. For this purpose, we tested two control strategies: open-loop and closed-loop control, and two stimulation paradigms: non-interleaved intramuscular stimulation (IM-S) and interleaved intraspinal microstimulation (ISMS). The experiments were conducted in anesthetized cats. Stimulation was applied to the muscles through IM-S electrodes implanted in the main knee and ankle extensor muscles, or to the spinal cord through ultra-fine ISMS wires implanted within the ventral horn of the lumbosacral enlargement. The cats were partially supported over parallel force plates and accelerometers were secured to the hindlimbs above and below the ankle joint. Ground reaction forces and knee and ankle joint angles were measured by the force plates and accelerometers, respectively. The closed-loop controller used these feedback signals to modulate the amplitude of stimulation applied to the extensor muscles. The open-loop controller applied constant levels of stimulation which were determined before the onset of each trial. The duration of standing achieved using closed-loop control of IM-S was significantly longer than that achieved with open-loop control (~2 times longer). The increase in the duration of standing corresponded with a decrease in the rate of force decay and a lower average injected current during closed-loop control. Standing was further improved with the use of ISMS. Closed-loop control of interleaved ISMS resulted in a period of standing > 3 times longer than the best trial generated using non-interleaved IM-S. There was also a significant improvement in the balance of force between the two hindlimbs. The results suggest that a system which uses closed-loop control in conjunction with interleaved ISMS could achieve prolonged FES standing in peop- - le with SCI.
  • Keywords
    biological techniques; biomechanics; bone; muscle; FES; SCI; accelerometer; anesthetized cat; ankle joint; closed-loop control; extensor muscle; functional electrical stimulation; ground reaction force; hindlimbs; intra spinal microstimulation; intramuscular stimulation; knee; lumbosacral enlargement; open-loop control; parallel force plate; prolonged functional standing; spinal cord injury; ventral horn; Adult cat; closed-loop control; feedback; functional electrical stimulation; open-loop control; spinal cord injury; Algorithms; Animals; Cats; Electric Stimulation Therapy; Feedback; Female; Male; Microelectrodes; Muscle Contraction; Muscle, Skeletal; Musculoskeletal Equilibrium; Posture; Spinal Cord; Spinal Cord Injuries; Therapy, Computer-Assisted; Treatment Outcome;
  • fLanguage
    English
  • Journal_Title
    Neural Systems and Rehabilitation Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1534-4320
  • Type

    jour

  • DOI
    10.1109/TNSRE.2007.897030
  • Filename
    4237175