Title :
Unambiguous measurement of volume flow using ultrasound
Author :
Hottinger, C.F. ; Meindl, J.D.
Author_Institution :
Stanford University, Stanford, Calif.
fDate :
6/1/1975 12:00:00 AM
Abstract :
A new modality for transcutaneous measurement of blood flow is described. Two ultrasonic pulsed Doppler techniques (Doppler imaging and average Doppler shift detection) are combined to measure net flow across an arbitrary sample plane. This estimate of volume flow is independent of lumen shape, orientation, and velocity profile.
Keywords :
Acoustic noise; Doppler shift; FETs; Fluid flow measurement; Rayleigh scattering; Shape measurement; Ultrasonic imaging; Ultrasonic transducers; Ultrasonic variables measurement; Volume measurement;
Journal_Title :
Proceedings of the IEEE
DOI :
10.1109/PROC.1975.9867