• DocumentCode
    935111
  • Title

    Background events in microchannel plates

  • Author

    Siegmund, O.H.W. ; Vallerga, J. ; Wargelin, B.

  • Author_Institution
    Space Sci. Lab., California Univ., Berkeley, CA, USA
  • Volume
    35
  • Issue
    1
  • fYear
    1988
  • Firstpage
    524
  • Lastpage
    528
  • Abstract
    Measurements have been made to assess the characteristics and origins of background events in microchannel plates (MCPs). An overall background rate of approximately 0.4 events cm/sup -3/-s/sup -1/ has been achieved consistently for MCPs that have been baked and scrubbed. The temperature and gain of the MCPs are found to have no significant effect on the background rate. Detection of 1.46-MeV gamma rays from the MCP glass confirms the presence of /sup 40/K, with a concentration of 0.0007%, in MCP glass. It is shown that beta decay from /sup 40/K is sufficient to cause the background rate and spectrum observed. Anticoincidence measurements indicate the background rate caused by cosmic ray interactions is small (<0.016 events cm/sup -2/ s/sup -1/).<>
  • Keywords
    electron multipliers; image intensifiers; position sensitive particle detectors; /sup 40/K; background events; beta decay; cosmic ray interactions; gain; microchannel plates; temperature; Event detection; Extraterrestrial measurements; Glass; Laboratories; Microchannel; Pulse amplifiers; Pulse measurements; Pulse shaping methods; Shape; Temperature;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.12778
  • Filename
    12778